Specifications affected by preamp – Agilent Technologies N9010A User Manual
Page 140
140
Chapter 13
Option P03, P07, P13, P26, P32 and P44 - Preamplifier
Specifications Affected by Preamp
Specifications Affected by Preamp
Specification Name
Information
Nominal Dynamic Range vs.
Offset Frequency vs. RBW
The graphic from the core specifications does not apply with
Preamp On.
Measurement Range
The measurement range depends on DANL.
See
“Amplitude Accuracy and Range” on page 33
Gain Compression
See specifications in this chapter.
Displayed Average Noise Level
See specifications in this chapter.
Displayed Average Noise Level
with Option MPB for Option 532
or 544
Performance is nominally 3 dB worse than without Option MPB.
Frequency Response
See specifications in this chapter.
Absolute Amplitude Accuracy
See
“Absolute Amplitude Accuracy” on page 37
specifications.
RF Input VSWR
See plot in this chapter.
Display Scale Fidelity
See
“Display Scale Fidelity” on page 41
of the core specifications.
Then, adjust the mixer levels given downward by the preamp gain
given in this chapter.
Second Harmonic Distortion
SHI with preamplifiers is not specified.
Third Order Intermodulation
Distortion
See specifications in this chapter.
Other Input Related Spurious
See
“Spurious Responses” on page 48
of the core specifications.
Preamp performance is not warranted but is nominally the same as
non-preamp performance.
Dynamic Range
See plot in this chapter.
Gain
See “Preamp” specifications in this chapter.
Noise Figure
See “Preamp” specifications in this chapter.