Chapter 1
Agilent EXA Signal Analyzer
Amplitude Accuracy and Range
f. Slope error will nominally be well within the range of
±0.0009 × (P1 − P2). P1 and P2 are defined in
footnote
g. A small additional error is possible. In FFT sweeps, this error is possible for spans under 4.01 kHz. For
non-FFT measurements, it is possible for RBWs of 3.9 kHz or less. The error is well within the range of
±0.0021 × (P1 - P2) subject to a maximum of ±0.005 dB. (The maximum dominates for all but very
small differences.) P1 and P2 are defined in footnote
Description
Specifications
Supplemental Information
Available Detectors
Normal, Peak, Sample, Negative Peak,
Average
Average detector works on RMS,
Voltage and Logarithmic scales