List sequence measurements – Agilent Technologies N9010A User Manual
Page 316
![background image](/manuals/35678/316/background.png)
316
Chapter 35
WLAN Measurement Application
List Sequence Measurements
List Sequence Measurements
1
1. Requires Option N9077A-5FP be installed and licensed.
Description
Specifications
Supplemental Information
Trtansmit Power
20 MHz Integration BW
Radio standard is:
802.11a/g/j/p (OFDM),
802.11g (DSSS-OFDM),
802.11n (20 MHz) or
802.11ac (20 MHz)
Center Frequency in 2.4 GHz Band
Minimum power at RF Input
–35 dBm (nominal)
Absolute Power Accuracy
a
(20 to 30°C)
a. Absolute power accuracy includes all error sources for in-band signals except mismatch errors and
repeatability due to incomplete averaging. It applies when the mixer level is high enough that the
measurement floor contribution is negligible.
±0.46 dB (nomonal)
Measurement floor
–72.7 dBm (typical)
Description
Specifications
Supplemental Information
Transmit Power
20 MHz Integration BW
Radio standard is:
802.11a/g/j/p (OFDM),
802.11n (20 MHz) or
802.11ac (20 MHz)
Center Frequency in 5.0 GHz Band
Minimum power at RF Input
–35 dBm (nominal)
Absolute Power Accuracy
a
(20 to 30°C)
a. Absolute power accuracy includes all error sources for in-band signals except mismatch errors and
repeatability due to incomplete averaging. It applies when the mixer level is high enough that the
measurement floor contribution is negligible.
±0.93 dB (nomonal)
Measurement floor
–72.7 dBm (typical)