Measurements, Channel power, Transmit on/off power – Agilent Technologies N9010A User Manual
Page 267: Channel power transmit on/off power
![background image](/manuals/35678/267/background.png)
Chapter 30
267
LTE Measurement Application
Measurements
Measurements
Description
Specifications
Supplemental Information
Channel Power
Minimum power at RF input
−50 dBm (nominal)
Absolute power accuracy
a
(20 to 30
°C, Atten = 10 dB)
a. Absolute power accuracy includes all error sources for in-band signals except mismatch errors and
repeatability due to incomplete averaging. It applies when the mixer level is high enough that the mea-
surement floor contribution is negligible.
±0.94 dB
95th Percentile
Absolute power accuracy
(20 to 30
°C, Atten = 10 dB)
±0.27 dB
Measurement floor
−75.7 dBm (nominal) in a 10 MHz
bandwidth
Description
Specifications
Supplemental Information
Transmit On/Off Power
This table applies only to the
N9082A measurement application.
Burst Type
Traffic, DwPTS, UpPTS, SRS,
PRACH
Transmit power
Min, Max, Mean, Off
Dynamic Range
a
a. This dynamic range expression is for the case of Information BW = 5 MHz; for other Info BW, the
dynamic range can be derived. The equation is:
Dynamic Range = Dynamic Range for 5 MHz – 10*log
10
(Info BW/5.0e6)
122.5 dB (nominal)
Average type
Off, RMS, Log
Measurement time
Up to 20 slots
Trigger source
External 1, External 2, Periodic, RF
Burst, IF Envelope