Frequency and time, Frequency range, Band – Agilent Technologies N9010A User Manual
Page 21: Frequency range band
Chapter 1
21
Agilent EXA Signal Analyzer
Frequency and Time
Frequency and Time
Description
Specifications
Supplemental Information
Frequency Range
Maximum Frequency
Option 503
3.6 GHz
Option 507
7 GHz
Option 513
13.6 GHz
Option 526
26.5 GHz
Option 532
32 GHz
Option 544
44 GHz
Preamp Option P03
3.6 GHz
Preamp Option P07
7 GHz
Preamp Option P13
13.6 GHz
Preamp Option P26
26.5 GHz
Preamp Option P32
32 GHz
Preamp Option P44
44 GHz
Minimum Frequency
Preamp
AC Coupled
a
a. AC Coupled only applicable to Freq Options 503, 507, 513, and 526.
DC Coupled
Off
10 MHz
10 Hz
On
10 MHz
100 kHz
Band
Harmonic
Mixing Mode
LO Multiple (N
b
)
b. N is the LO multiplication factor. For negative mixing modes (as indicated by the “
−” in the “Har-
monic Mixing Mode” column), the desired 1st LO harmonic is higher than the tuned frequency by the
1st IF (5.1225 GHz for band 0, 322.5 MHz for all other bands).
Band Overlaps
c
0 (10 Hz to 3.6 GHz)
1
−
1
Options 503, 507, 513, 526, 532,
544
1 (3.5 GHz to 7 GHz)
1
−
1
Option 507
1 (3.5 GHz to 8.4 GHz)
1
−
1
Options 513, 526, 532, 544
2 (8.3 GHz to 13.6 GHz)
1
−
2
Options 513, 526, 532, 544
3 (13.5 to 17.1 GHz)
2
−
2
Options 526, 532, 544
4 (17.0 to 26.5 GHz)
2
−
4
Options 526, 532, 544
5 (26.4 GHz to 32 GHz)
2
−
4
Option 532
5 (26.4 GHz to 34.5 GHz)
2
−
4
Option 544
6 (34.4 GHz to 44 GHz)
4
−
8
Option 544