beautypg.com

Frequency and time, Frequency range, Band – Agilent Technologies N9010A User Manual

Page 21: Frequency range band

background image

Chapter 1

21

Agilent EXA Signal Analyzer

Frequency and Time

Frequency and Time

Description

Specifications

Supplemental Information

Frequency Range

Maximum Frequency

Option 503

3.6 GHz

Option 507

7 GHz

Option 513

13.6 GHz

Option 526

26.5 GHz

Option 532

32 GHz

Option 544

44 GHz

Preamp Option P03

3.6 GHz

Preamp Option P07

7 GHz

Preamp Option P13

13.6 GHz

Preamp Option P26

26.5 GHz

Preamp Option P32

32 GHz

Preamp Option P44

44 GHz

Minimum Frequency

Preamp

AC Coupled

a

a. AC Coupled only applicable to Freq Options 503, 507, 513, and 526.

DC Coupled

Off

10 MHz

10 Hz

On

10 MHz

100 kHz

Band

Harmonic
Mixing Mode

LO Multiple (N

b

)

b. N is the LO multiplication factor. For negative mixing modes (as indicated by the “

−” in the “Har-

monic Mixing Mode” column), the desired 1st LO harmonic is higher than the tuned frequency by the
1st IF (5.1225 GHz for band 0, 322.5 MHz for all other bands).

Band Overlaps

c

0 (10 Hz to 3.6 GHz)

1

1

Options 503, 507, 513, 526, 532,
544

1 (3.5 GHz to 7 GHz)

1

1

Option 507

1 (3.5 GHz to 8.4 GHz)

1

1

Options 513, 526, 532, 544

2 (8.3 GHz to 13.6 GHz)

1

2

Options 513, 526, 532, 544

3 (13.5 to 17.1 GHz)

2

2

Options 526, 532, 544

4 (17.0 to 26.5 GHz)

2

4

Options 526, 532, 544

5 (26.4 GHz to 32 GHz)

2

4

Option 532

5 (26.4 GHz to 34.5 GHz)

2

4

Option 544

6 (34.4 GHz to 44 GHz)

4

8

Option 544