If phase linearity, Absolute amplitude accuracy, If phase linearity absolute amplitude accuracy – Agilent Technologies N9010A User Manual
Page 37
![background image](/manuals/35678/37/background.png)
Chapter 1
37
Agilent EXA Signal Analyzer
Amplitude Accuracy and Range
Description
Specifications Supplemental Information
IF Phase Linearity
Deviation from mean phase linearity
Modes above 18 GHz
a
a. Signal frequencies between 18 and 26.5 GHz are prone to additional response errors due to modes in the
Type-N connector used with frequency Option 526. With the use Type-N to APC 3.5 mm adapter part
number 1250-1744, there are nominally six such modes. These modes cause nominally up to
−0.35 dB
amplitude change, with phase errors of nominally up to
±1.2
°
.
Center Freq (GHz)
Span
(MHz)
Preselector
Peak-to-peak
(nominal
)
RMS (nominal)
b
b. The listed performance is the standard deviation of the phase deviation relative to the mean phase devi-
ation from a linear phase condition, where the rms is computed across the span shown and over the
range of center frequencies shown.
≥0.02, <3.6
≤10
n/a
0.4°
0.1°
≥3.6,
≤10
Off
c
c. Option MPB is installed and enabled.
0.4°
0.1°
≥3.6 (Option ≤ 526) ≤10
On
1.0°
0.2°
Description
Specifications
Supplemental Information
Absolute Amplitude Accuracy
At 50 MHz
a
20 to 30
°C
Full temperature range
±0.40 dB
±0.43 dB
±0.15 dB
(95th percentile)
At all frequencies
20 to 30
°C
Full temperature range
±(0.40 dB + frequency response)
±(0.43 dB + frequency response)
95th Percentile Absolute
Amplitude Accuracy
b
±0.27 dB
(Wide range of signal levels,
RBWs, RLs, etc.,
0.01 to 3.6 GHz,
Atten = 10 dB)
Amplitude Reference Accuracy
±0.05 dB (nominal)
Preamp On
c
±(0.39 dB + frequency
response) (nominal)