105

Option B40 - 40 MHz Analysis Bandwidth

Data Acquisition

Data Acquisition

Description

Specifications

Supplemental
Information

Time Record Length

IQ Analyzer

4,000,000 IQ sample pairs

Advanced Tools

Data Packing

89600 VSA software or
N9064A

a

VXA

a. In software versions prior to A.06.00, the VXA measurement application product number was

89601X. Software versions A.06.00 and beyond have renamed 89601X to N9064A.

32-bit 64-bit

Length (IQ sample pairs) 536 MSa (2

29

Sa)

268 MSa (2

28

Sa)

2 GB total memory

Length (time units)

Samples/(Span

× 1.28)

Sample Rate

At ADC

200 MSa/s

IQ Pairs

Span dependent

ADC Resolution

12 bits

Capture Time [Plot]

NOTE

This plot is based on the full access to the 2 GB deep capture memory which
requires either the 89600 VSA or the N9064A VXA