Agilent Technologies N9010A User Manual
Page 136
136
Chapter 12
Option MPB - Microwave Preselector Bypass
Specifications Affected by Microwave Preselector Bypass
Specifications Affected by Microwave Preselector Bypass
Specification Name
Information
Displayed Average Noise Level
For analyzers with frequency Option 526 (26.5 GHz) or lower:
Performance is not identical, but nominally the same, as
without Option MPB.
For analyzers with frequency option higher than Option 526
(26.5 GHz): Performance is nominally 3 dB better than without
Option MPB.
IF Frequency Response and IF Phase
Linearity
See
“IF Frequency Response” on page 36
for the standard 10 MHz analysis
bandwidth; also, see the associated "Analysis Bandwidth"
chapter for any optional bandwidths.
Frequency Response
See specifications in this chapter.
VSWR
The magnitude of the mismatch over the range of frequencies
will be very similar between MPB and non-MPB operation, but
the details, such as the frequencies of the peaks and valleys,
will shift.
Additional Spurious Responses
“Spurious Responses” on page 48
of the core
specifications,
“Additional Spurious Responses” on page 138
of this chapter also apply.