beautypg.com

Agilent Technologies N9010A User Manual

Page 136

background image

136

Chapter 12

Option MPB - Microwave Preselector Bypass

Specifications Affected by Microwave Preselector Bypass

Specifications Affected by Microwave Preselector Bypass

Specification Name

Information

Displayed Average Noise Level

For analyzers with frequency Option 526 (26.5 GHz) or lower:
Performance is not identical, but nominally the same, as
without Option MPB.

For analyzers with frequency option higher than Option 526
(26.5 GHz): Performance is nominally 3 dB better than without
Option MPB.

IF Frequency Response and IF Phase
Linearity

See

“IF Frequency Response” on page 36

and

“IF Phase

Linearity” on page 37

for the standard 10 MHz analysis

bandwidth; also, see the associated "Analysis Bandwidth"
chapter for any optional bandwidths.

Frequency Response

See specifications in this chapter.

VSWR

The magnitude of the mismatch over the range of frequencies
will be very similar between MPB and non-MPB operation, but
the details, such as the frequencies of the peaks and valleys,
will shift.

Additional Spurious Responses

In addition to the

“Spurious Responses” on page 48

of the core

specifications,

“Additional Spurious Responses” on page 138

of this chapter also apply.