Chapter 21
Bluetooth Measurement Application
Basic Rate Measurements
Description
Specifications
Supplemental Information
Initial Carrier Frequency
Tolerance
This measurement is a Transmit
Analysis measurement and
supports average and peak power
in conformance with Bluetooth
RF test specification
2.1.E.0.5.1.10.
Packet Type
DH1, DH3, DH5, HV3
Payload
PRBS9, BS00, BSFF, BS0F,
BS55
Synchronization
Preamble
Trigger
External, RF Burst, Periodic
Timer, Free Run, Video
RF input level range
a
+30 dBm to –70 dBm
Measurement range
Nominal channel freq ± 100 kHz
(nominal)
Measurement Accuracy
b
±100 Hz + tfa
c
(nominal)
a. When the input signal level is lower than –40 dBm, the analyzer’s preamp should be turned on and the
attenuator set to 0 dB.
b. Example, using 1 ppm as frequency reference accuracy of the analyzer, at frequency of 2.402 GHz,
frequency accuracy would be in the range of ±(2.402 GHz
× 1 ppm) Hz ± 100 Hz = ±2402 Hz ± 100 Hz
= ±2502 Hz.