Option P03, P07, P13, P26, P32 and P44 - Preamplifier
Other Preamp Specifications
20 to 26.5. GHz
x
−150 dBm
−147 dBm
−156 dBm
20 to 26.5 GHz
x
−157 dBm
−155 dBm
−159 dBm
20 to 26.5 GHz
x
−158 dBm
−156 dBm
−160 dBm
26.5 to 32 GHz
x
−155 dBm
−152 dBm
−158 dBm
26.5 to 32 GHz
x
−156 dBm
−153 dBm
−159 dBm
Option P44
32 to 34 GHz
x
−155 dBm
−152 dBm
−158 dBm
32 to 34 GHz
x
−156 dBm
−153 dBm
−159 dBm
33.9 to 40 GHz
x
−152 dBm
−148 dBm
−154 dBm
33.9 to 40 GHz
x
−153 dBm
−150 dBm
−155 dBm
40 to 44 GHz
x
−148 dBm
−144 dBm
−152 dBm
40 to 44 GHz
x
−149 dBm
−146 dBm
−153 dBm
a. DANL is measured in a 1 kHz RBW and normalized to the narrowest available RBW, because the noise
figure does not depend on RBW and 1 kHz measurements are faster.
b. Specifications apply only when the Phase Noise Optimization control is set to “Best Wide-offset Phase
Noise.”
Description
Specifications
Supplemental Information
Frequency Response
− Preamp
On
(Options P03, P07, P13, P26, P32,
P44)
(Maximum error relative to
reference condition (50 MHz)
Input attenuation 0 dB
Swept operation
a
)
100 kHz to 3.6 GHz
b
±0.28 dB (nominal)
3.5 to 8.4 GHz
±0.67 dB (nominal)
8.3 to 26.5 GHz
±0.8 dB (nominal)
26.5 to 44 GHz
±0.8 dB (nominal)
a. For Sweep Type = FFT, add the RF flatness errors of this table to the IF Frequency Response errors. An
additional error source, the error in switching between swept and FFT sweep types, is nominally
±0.01
dB and is included within the “Absolute Amplitude Error” specifications.
b. Electronic attenuator (Option EA3) may not be used with preamp on.
Description
Specifications
Supplemental