beautypg.com

Frequency response - preamp on, Frequency response – Agilent Technologies N9010A User Manual

Page 144

background image

144

Chapter 13

Option P03, P07, P13, P26, P32 and P44 - Preamplifier

Other Preamp Specifications

20 to 26.5. GHz

x

−150 dBm

−147 dBm

−156 dBm

20 to 26.5 GHz

x

−157 dBm

−155 dBm

−159 dBm

20 to 26.5 GHz

x

−158 dBm

−156 dBm

−160 dBm

26.5 to 32 GHz

x

−155 dBm

−152 dBm

−158 dBm

26.5 to 32 GHz

x

−156 dBm

−153 dBm

−159 dBm

Option P44

32 to 34 GHz

x

−155 dBm

−152 dBm

−158 dBm

32 to 34 GHz

x

−156 dBm

−153 dBm

−159 dBm

33.9 to 40 GHz

x

−152 dBm

−148 dBm

−154 dBm

33.9 to 40 GHz

x

−153 dBm

−150 dBm

−155 dBm

40 to 44 GHz

x

−148 dBm

−144 dBm

−152 dBm

40 to 44 GHz

x

−149 dBm

−146 dBm

−153 dBm

a. DANL is measured in a 1 kHz RBW and normalized to the narrowest available RBW, because the noise

figure does not depend on RBW and 1 kHz measurements are faster.

b. Specifications apply only when the Phase Noise Optimization control is set to “Best Wide-offset Phase

Noise.”

Description

Specifications

Supplemental Information

Frequency Response

Preamp

On

(Options P03, P07, P13, P26, P32,
P44)

(Maximum error relative to

reference condition (50 MHz)

Input attenuation 0 dB

Swept operation

a

)

100 kHz to 3.6 GHz

b

±0.28 dB (nominal)

3.5 to 8.4 GHz

±0.67 dB (nominal)

8.3 to 26.5 GHz

±0.8 dB (nominal)

26.5 to 44 GHz

±0.8 dB (nominal)

a. For Sweep Type = FFT, add the RF flatness errors of this table to the IF Frequency Response errors. An

additional error source, the error in switching between swept and FFT sweep types, is nominally

±0.01

dB and is included within the “Absolute Amplitude Error” specifications.

b. Electronic attenuator (Option EA3) may not be used with preamp on.

Description

Specifications

Supplemental