Other microwave preselector bypass specifications – Agilent Technologies N9010A User Manual
Page 137
![background image](/manuals/35678/137/background.png)
Chapter 12
137
Option MPB - Microwave Preselector Bypass
Other Microwave Preselector Bypass Specifications
Other Microwave Preselector Bypass Specifications
Description
Specifications
Supplemental Information
Frequency Response
(Maximum error relative to
reference condition (50 MHz)
Swept operation
a
, Attenuation
10 dB)
a. For Sweep Type = FFT, add the RF flatness errors of this table to the IF Frequency Response errors. An
additional error source, the error in switching between swept and FFT sweep types, is nominally
±0.01 dB and is included within the “Absolute Amplitude Error” specifications.
Refer to the footnote for
.
Modes above 18 GHz
b
b. Signal frequencies between 18 and 26.5 GHz are prone to additional response errors due to modes in
the Type-N connector used with frequency Option 526. With the use Type-N to APC 3.5 mm adapter
part number 1250-1744, there are nominally six such modes. The effect of these modes with this con-
nector are included within these specifications.
20 to 30
°C
Full Range
95
th
Percentile (
≈2σ)
3.5 to 8.4 GHz
±0.9 dB
±1.5 dB
±0.42 dB
8.3 to 13.6 GHz
±1.0 dB
±2.0 dB
±0.50 dB
13.5 to 17.1 GHz
±1.3 dB
±2.0 dB
±0.50 dB
17.0 to 22.0 GHz
±1.3 dB
±2.0 dB
±0.53 dB
22.0 to 26.5 GHz
±2.0 dB
±2.8 dB
±0.66 dB
26.4 to 34.5 GHz
±2.0 dB
±3.0 dB
±0.80 dB
34.4 to 44 GHz
±3.1 dB
±4.8 dB
±1.21 dB