Low Energy Measurements

Description

Specifications

Supplemental Information

Initial Carrier Frequency
Tolerance

This measurement is a Transmit
Analysis measurement and is in
conformance with Bluetooth RF
test specification
LE.RF-PHY.TS/0.7d2.6.2.4.

Packet Type

Reference type

Payload

PRBS9, BS00, BSFF, BS0F,
BS55

Synchronization

Preamble

Trigger

External, RF Burst, Periodic
Timer, Free Run, Video

RF input level range

a

+30 dBm to –70 dBm

Measurement range

Nominal channel freq ± 100 kHz
(nominal)

Measurement Accuracy

b

±100 Hz + tfa

c

(nominal)

a. When the input signal level is lower than –40 dBm, the analyzer’s preamp should be turned on and the

attenuator set to 0 dB.

b. Example, using 1 ppm as frequency reference accuracy of the analyzer, at frequency of 2.402 GHz,

frequency accuracy would be in the range of ±(2.402 GHz

× 1 ppm) Hz ± 100 Hz = ±2402 Hz ± 100 Hz

= ±2502 Hz.

c. tfa = transmitter frequency × frequency reference accuracy.