beautypg.com

Adjacent channel power, Spectrum emission mask, Adjacent channel power spectrum emission mask – Agilent Technologies N9010A User Manual

Page 257

background image

Chapter 29

257

ISDB-T Measurement Application

Measurements

Description

Specifications

Supplemental Information

Adjacent Channel Power

Minimum power at RF Input

–36 dBm (nominal)

ACPR Accuracy

a

(5.60 MHz noise bandwidth

method = IBW,

Offset Freq = 6 MHz)

±0.81 dB

At ACPR –45 dBc with
optimum mixer level

b

a. The accuracy of the Adjacent Channel Power Ratio will depend on the mixer drive level and whether

the distortion products from the analyzer are coherent with those in the UUT. These specifications
apply even in the worst case condition of coherent analyzer and UUT distortion products. For ACPR
levels other than those in this specifications table, the optimum mixer drive level for accuracy is
approximately –37 dBm – (ACPR/3), where the ACPR is given in (negative) decibels.

b. To meet this specified accuracy when measuring transmitter at –45 dBc ACPR, the mixer level (ML)

must be optimized for accuracy. This optimum mixer level is –20 dBm, so the input attenuation must
be set as close as possible to the average input power. For example, if the average input power is
–3 dBm, set the attenuation to 17 dB. Note that if the mixer level is set to optimize dynamic range
instead of accuracy, accuracy errors are nominally doubled.

Description

Specifications

Supplemental Information

Spectrum Emission Mask


(5.60 MHz Integration BW

RBW = 10.0 kHz)

Limit Type

Manual

JEITA (ARIB-B31) according to

P

≤ 0.025 W;

0.025 W

< P ≤ 0.25 W;

0.25 W

< P ≤ 2.5 W;

P

> 2.5 W

(P is the channel power)

ABNT Non-Critical

ABNT Sub-Critical

ABNT Critical

ISDB-T

SB

3.0 MHz Offset

Dynamic Range, relative

ab

82.3 dB

89.5 dB (typical)