Noise figure uncertainty calculator – Agilent Technologies N9010A User Manual
Page 172
![background image](/manuals/35678/172/background.png)
172
Chapter 17
Noise Figure Measurement Application
General Specifications
Description
Specifications
Supplemental Information
Noise Figure Uncertainty Calculator
a
a. The Noise Figure Uncertainty Calculator requires the parameters shown in order to calculate the total
uncertainty of a Noise Figure measurement.
Instrument Noise Figure Uncertainty
See the Noise Figure
table earlier in this
chapter
Instrument Gain Uncertainty
See the Gain table
earlier in this chapter
Instrument Noise Figure
See graphs of “Nominal Instrument
Noise Figure”; Noise Figure is
DANL + 176.24 dB (nominal)
b
Note on DC coupling
cd
b. Nominally, the noise figure of the spectrum analyzer is given by
NF = D
− (K − L + N + B)
where D is the DANL (displayed average noise level) specification,
K is kTB (
−173.98 dBm in a 1 Hz bandwidth at 290 K)
L is 2.51 dB (the effect of log averaging used in DANL verifications)
N is 0.24 dB (the ratio of the noise bandwidth of the RBW filter with which DANL is
specified to an ideal noise bandwidth)
B is ten times the base-10 logarithm of the RBW (in hertz) in which the DANL is
specified. B is 0 dB for the 1 Hz RBW.
The actual NF will vary from the nominal due to frequency response errors.
c. The effect of AC coupling is negligible for frequencies above 40 MHz. Below 40 MHz, DC coupling is
recommended for the best measurements.
d. The instrument NF nominally degrades by 0.2 dB at 30 MHz and 1 dB at 10 MHz with AC coupling.
Instrument Input Match
See graphs: Nominal VSWR
Note on DC coupling