Amplitude, Emi average detector, Quasi-peak detector – Agilent Technologies N9010A User Manual
Page 124: Rms average detector
![background image](/manuals/35678/124/background.png)
124
Chapter 9
Option EMC - Precompliance EMI Features
Amplitude
Amplitude
Description
Specifications
Supplemental Information
EMI Average Detector
Used for CISPR-compliant average
measurements and, with 1 MHz
RBW, for frequencies above 1 GHz
Default Average Type
All filtering is done on the linear
(voltage) scale even when the display
scale is log.
Quasi-Peak Detector
Used with CISPR-compliant RBWs,
for frequencies
≤ 1 GHz
Absolute Amplitude Accuracy for
reference spectral intensities
Meets CISPR standards
a
a. CISPR 16-1-1 (2010)
Relative amplitude accuracy versus pulse
repetition rate
Meets CISPR standards
Quasi-Peak to average response ratio
Meets CISPR standards
Dynamic range
Pulse repetition rates
≥ 20 Hz
Meets CISPR standards
Pulse repetition rates
≤ 10 Hz
Does not meet CISPR standards in
some cases with DC pulse excitation.
RMS Average Detector
Meets CISPR standards