Power sweep range, Measurement time, Power sweep range measurement time – Agilent Technologies N9010A User Manual

Page 128

background image

128

Chapter 10

Option ESC - External Source Control

General Specifications

Description

Specification

Supplemental Information

Power Sweep Range

Limited by source amplitude range

Description

Specification

Supplemental Information

Measurement Time

Nominal

a

a. These measurement times were observed with a span of 100 MHz, RBW of 20 kHz, and the point trig-

gering method being set to Ext Trigger1. The measurement times will not change significantly with
span when the RBW is automatically selected. If the RBW is decreased, the sweep time increase would
be approximately 23.8 times Npoints/RBW.

RF MXG (N5181A/N5182A)

b

b. Based on MXG firmware version A.01.80 and Option UNZ installed.

Option 503, 507, 513, 526, 532, 544

Band 0

Band 1

201 Sweep points (default setting)

450 ms

1.1 s

601 Sweep points

1.25 s

3.7 s

μW MXG (N5183A)

b

Option 503, 507, 513, 526

Band 0

Band 1

201 Sweep points (default setting)

450 ms

1.2 s

2.4 s

601 Sweep points

1.2 s

3.7 s

6.9 s

Option 532, 544

201 Sweep points (default setting)

450 ms

6.5 s

6.6 s

601 Sweep points

1.2 s

19 s

19.1 s

PSG (E8257D)/(E8267D)

c

c. Based on PSG firmware version C.06.15 and Option UNZ installed.

Option 503, 507, 513, 526

Band 0

Band 1

201 Sweep points (default setting)

2.2 s

2.2 s

2.5 s

601 Sweep points

6.1 s

6.5 s

7.1 s

Option 532, 544

201 Sweep points (default setting)

2.2 s

6.6 s

6.6 s

601 Sweep points

6.1 s

19.5 s

19.1 s