Chapter 10
Option ESC - External Source Control
General Specifications
Description
Specification
Supplemental Information
Power Sweep Range
Limited by source amplitude range
Description
Specification
Supplemental Information
Measurement Time
Nominal
a
a. These measurement times were observed with a span of 100 MHz, RBW of 20 kHz, and the point trig-
gering method being set to Ext Trigger1. The measurement times will not change significantly with
span when the RBW is automatically selected. If the RBW is decreased, the sweep time increase would
be approximately 23.8 times Npoints/RBW.
RF MXG (N5181A/N5182A)
b
b. Based on MXG firmware version A.01.80 and Option UNZ installed.
Option 503, 507, 513, 526, 532, 544
Band 0
Band 1
201 Sweep points (default setting)
450 ms
1.1 s
601 Sweep points
1.25 s
3.7 s
μW MXG (N5183A)
Option 503, 507, 513, 526
Band 0
Band 1
201 Sweep points (default setting)
450 ms
1.2 s
2.4 s
601 Sweep points
1.2 s
3.7 s
6.9 s
Option 532, 544
201 Sweep points (default setting)
450 ms
6.5 s
6.6 s
601 Sweep points
1.2 s
19 s
19.1 s
PSG (E8257D)/(E8267D)
c
c. Based on PSG firmware version C.06.15 and Option UNZ installed.
Option 503, 507, 513, 526
Band 0
Band 1
201 Sweep points (default setting)
2.2 s
2.2 s
2.5 s
601 Sweep points
6.1 s
6.5 s
7.1 s
Option 532, 544
201 Sweep points (default setting)
2.2 s
6.6 s
6.6 s
601 Sweep points
6.1 s
19.5 s
19.1 s