Amplitude accuracy and range, Measurement range, Maximum safe input level – Agilent Technologies N9010A User Manual
Page 33: Display range
![background image](/manuals/35678/33/background.png)
Chapter 1
33
Agilent EXA Signal Analyzer
Amplitude Accuracy and Range
Amplitude Accuracy and Range
Description
Specifications
Supplemental
Information
Measurement Range
Preamp Off
Displayed Average Noise Level to +23 dBm
Preamp On
Displayed Average Noise Level to +23 dBm
Option P03, P07, P13, P26,
P32, P44
Input Attenuation Range
Standard
0 to 60 dB, in 10 dB steps
With Option FSA
0 to 60 dB, in 2 dB steps
Description
Specifications
Supplemental Information
Maximum Safe Input Level
Applies with or without preamp
(Option P03, P07, P13, P26, P32, P44)
Average Total Power
+30 dBm (1 W)
Peak Pulse Power
(≤10 μs pulse width,
≤1% duty cycle,
input attenuation
≥ 30 dB)
+50 dBm (100 W)
DC voltage
DC Coupled
±0.2 Vdc
AC Coupled
±100 Vdc
Description
Specifications
Supplemental
Information
Display Range
Log Scale
Ten divisions displayed;
0.1 to 1.0 dB/division in 0.1 dB steps, and
1 to 20 dB/division in 1 dB steps
Linear Scale
Ten divisions