Amplitude Accuracy and Range

Amplitude Accuracy and Range

Description

Specifications

Supplemental
Information

Measurement Range

Preamp Off

Displayed Average Noise Level to +23 dBm

Preamp On

Displayed Average Noise Level to +23 dBm

Option P03, P07, P13, P26,
P32, P44

Input Attenuation Range

Standard

0 to 60 dB, in 10 dB steps

With Option FSA

0 to 60 dB, in 2 dB steps

Description

Specifications

Supplemental Information

Maximum Safe Input Level

Applies with or without preamp
(Option P03, P07, P13, P26, P32, P44)

Average Total Power

+30 dBm (1 W)

Peak Pulse Power

(≤10 μs pulse width,

≤1% duty cycle,
input attenuation

≥ 30 dB)

+50 dBm (100 W)

DC voltage

DC Coupled

±0.2 Vdc

AC Coupled

±100 Vdc

Description

Specifications

Supplemental
Information

Display Range

Log Scale

Ten divisions displayed;
0.1 to 1.0 dB/division in 0.1 dB steps, and
1 to 20 dB/division in 1 dB steps

Linear Scale