beautypg.com

Input attenuation switching uncertainty, Rf input vswr, Resolution bandwidth switching uncertainty – Agilent Technologies N9010A User Manual

Page 39

background image

Chapter 1

39

Agilent EXA Signal Analyzer

Amplitude Accuracy and Range

Description

Specifications

Supplemental Information

Input Attenuation Switching Uncertainty

Refer to the footnote for

Band Overlaps on page 21

50 MHz (reference frequency)

±0.20 dB

±0.08 dB (typical)

Attenuation > 2 dB, preamp off

(Relative to 10 dB (reference setting))

9 kHz to 3.6 GHz

±0.3 dB (nominal)

3.5 to 7.0 GHz

±0.5 dB (nominal)

7.0 to 13.6 GHz

±0.7 dB (nominal)

13.5 to 26.5 GHz

±0.7 dB (nominal)

26.5 to 44 GHz

±1.0 dB (nominal)

Description

Specifications Supplemental Information

RF Input VSWR

Nominal

a

a. The nominal SWR stated is at the worst case RF frequency in three representative instruments.

at tuned frequency, DC Coupled

10 dB attenuation, 50 MHz

1.07:1

Input Attenuation

Frequency

0 dB

10 dB

Option

526

10 MHz to 3.6 GHz

<2.2:1

<1.2:1

3.6 to 26.5 GHz

<1.8:1

Option

>526

10 MHz to 3.6 GHz

<2.2:1

<1.2:1

3.6 to 26.5 GHz

<1.5:1

26.5 to 44 GHz

<1.8:1

RF calibrator (e.g. 50 MHz) is On

Open input

Alignments running

Open input for some, unless "All but RF" is
selected

Preselector Centering

Open input

Description

Specifications

Supplemental Information

Resolution Bandwidth Switching Uncertainty

Relative to reference BW of
30 kHz

1.0 Hz to 3 MHz RBW

±0.10 dB

Manually selected wide RBWs: 4, 5, 6, 8 MHz

±1.0 dB