Input attenuation switching uncertainty, Rf input vswr, Resolution bandwidth switching uncertainty – Agilent Technologies N9010A User Manual
Page 39
Chapter 1
39
Agilent EXA Signal Analyzer
Amplitude Accuracy and Range
Description
Specifications
Supplemental Information
Input Attenuation Switching Uncertainty
Refer to the footnote for
50 MHz (reference frequency)
±0.20 dB
±0.08 dB (typical)
Attenuation > 2 dB, preamp off
(Relative to 10 dB (reference setting))
9 kHz to 3.6 GHz
±0.3 dB (nominal)
3.5 to 7.0 GHz
±0.5 dB (nominal)
7.0 to 13.6 GHz
±0.7 dB (nominal)
13.5 to 26.5 GHz
±0.7 dB (nominal)
26.5 to 44 GHz
±1.0 dB (nominal)
Description
Specifications Supplemental Information
RF Input VSWR
Nominal
a
a. The nominal SWR stated is at the worst case RF frequency in three representative instruments.
at tuned frequency, DC Coupled
10 dB attenuation, 50 MHz
1.07:1
Input Attenuation
Frequency
0 dB
≥10 dB
Option
≤526
10 MHz to 3.6 GHz
<2.2:1
<1.2:1
3.6 to 26.5 GHz
<1.8:1
Option
>526
10 MHz to 3.6 GHz
<2.2:1
<1.2:1
3.6 to 26.5 GHz
<1.5:1
26.5 to 44 GHz
<1.8:1
RF calibrator (e.g. 50 MHz) is On
Open input
Alignments running
Open input for some, unless "All but RF" is
selected
Preselector Centering
Open input
Description
Specifications
Supplemental Information
Resolution Bandwidth Switching Uncertainty
Relative to reference BW of
30 kHz
1.0 Hz to 3 MHz RBW
±0.10 dB
Manually selected wide RBWs: 4, 5, 6, 8 MHz
±1.0 dB