Low energy measurements, Output power – Agilent Technologies N9010A User Manual

Page 196

background image

196

Chapter 21

Bluetooth Measurement Application

Low Energy Measurements

Low Energy Measurements

Description

Specifications

Supplemental Information

Output Power

This measurement is a Transmit
Analysis measurement and
supports average and peak power
in conformance with Bluetooth
RF test specification
LE.RF-PHY.TS/0.7d2.6.2.1.

Packet Type

Reference type

Payload

PRBS9, BS00, BSFF, BS0F,
BS55

Synchronization

RF Burst or Preamble

Trigger

External, RF Burst, Periodic
Timer, Free Run, Video

Supported measurements

Average Power, Peak Power

Range

a

a. When the input signal level is lower than –40 dBm, the analyzer’s preamp should be turned on and the

attenuator set to 0 dB.

+30 dBm to –70 dBm

Absolute Power Accuracy

b

(20 to 30°C, Atten = 10 dB)

b. Absolute power accuracy includes all error sources for in-band signals except mismatch errors and

repeatability due to incomplete averaging. It applies when the mixer level is high enough that measure-
ment floor contribution is negligible.

±0.29 dB (95th percentile)

Measurement floor

–70 dBm (nominal)