Low energy measurements, Output power – Agilent Technologies N9010A User Manual
Page 196
196
Chapter 21
Bluetooth Measurement Application
Low Energy Measurements
Low Energy Measurements
Description
Specifications
Supplemental Information
Output Power
This measurement is a Transmit
Analysis measurement and
supports average and peak power
in conformance with Bluetooth
RF test specification
LE.RF-PHY.TS/0.7d2.6.2.1.
Packet Type
Reference type
Payload
PRBS9, BS00, BSFF, BS0F,
BS55
Synchronization
RF Burst or Preamble
Trigger
External, RF Burst, Periodic
Timer, Free Run, Video
Supported measurements
Average Power, Peak Power
Range
a
a. When the input signal level is lower than –40 dBm, the analyzer’s preamp should be turned on and the
attenuator set to 0 dB.
+30 dBm to –70 dBm
Absolute Power Accuracy
b
(20 to 30°C, Atten = 10 dB)
b. Absolute power accuracy includes all error sources for in-band signals except mismatch errors and
repeatability due to incomplete averaging. It applies when the mixer level is high enough that measure-
ment floor contribution is negligible.
±0.29 dB (95th percentile)
Measurement floor
–70 dBm (nominal)