Input, Range, Adc overload – Agilent Technologies N9010A User Manual
Page 85: Input range adc overload
![background image](/manuals/35678/85/background.png)
Chapter 3
85
VXA Vector Signal and WLAN Modulation Analysis Application
Vector Signal Analysis Performance (N9064A-1FP/1TP)
Input
Description
Specs & Nominals
Supplemental
Information
Range
Full Scale, combines
attenuator setting and
ADC gain
standard
−20 dBm to 20 dBm, 10 dB steps
Option FSA or EA3
−20 dBm to 22 dBm, 2 dB steps
Option P03, P07, P13, P26,
P32, P44 with neither FSA
or EA3
−40 dBm to 20 dBm, 10 dB steps, up to 3.6 GHz
Options P03, P07, P13, P26,
P32, P44 and either FSA or
EA3
−40 dBm to 22 dBm, 2 dB steps, up to 3.6 GHz
Option P07, P13, P26, P32,
or P44
−40 dBm to 20 dBm, 10 dB steps, above 3.6 GHz
ADC overload
+2 dBFS