beautypg.com

Rear panel – Agilent Technologies N9010A User Manual

Page 72

background image

72

Chapter 1

Agilent EXA Signal Analyzer

Inputs/Outputs

Rear Panel

Description

Specifications

Supplemental Information

10 MHz Out

Connector

BNC female

Impedance

50

Ω (nominal)

Output Amplitude

≥0 dBm (nominal)

Output Configuration

AC coupled, sinusoidal

Frequency

10 MHz

×

(1 + frequency reference accuracy)

Description

Specifications

Supplemental Information

Ext Ref In

Connector

BNC female

Note: Analyzer noise sidebands and
spurious response performance may be
affected by the quality of the external
reference used. See footnote

c

in the

Phase Noise specifications within the
Dynamic Range section on

page 53

.

Impedance

50

Ω (nominal)

Input Amplitude Range

sine wave

square wave

−5 to +10 dBm (nominal)
0.2 to 1.5 V peak-to-peak (nominal)

Input Frequency

10 MHz (nominal)

Lock range

±5 × 10

−6

of ideal external

reference input frequency

Description

Specifications

Supplemental Information

Sync

Reserved for future use

Connector

BNC female

Description

Specifications

Supplemental Information

Trigger Inputs

(Trigger 1 In, Trigger 2 In)

Either trigger source may be selected

Connector

BNC female

Impedance

10 k

Ω (nominal)

Trigger Level Range

−5 to +5 V

1.5 V (TTL) factory preset