Rear panel – Agilent Technologies N9010A User Manual
Page 72
72
Chapter 1
Agilent EXA Signal Analyzer
Inputs/Outputs
Rear Panel
Description
Specifications
Supplemental Information
10 MHz Out
Connector
BNC female
Impedance
50
Ω (nominal)
Output Amplitude
≥0 dBm (nominal)
Output Configuration
AC coupled, sinusoidal
Frequency
10 MHz
×
(1 + frequency reference accuracy)
Description
Specifications
Supplemental Information
Ext Ref In
Connector
BNC female
Note: Analyzer noise sidebands and
spurious response performance may be
affected by the quality of the external
reference used. See footnote
in the
Phase Noise specifications within the
Dynamic Range section on
Impedance
50
Ω (nominal)
Input Amplitude Range
sine wave
square wave
−5 to +10 dBm (nominal)
0.2 to 1.5 V peak-to-peak (nominal)
Input Frequency
10 MHz (nominal)
Lock range
±5 × 10
−6
of ideal external
reference input frequency
Description
Specifications
Supplemental Information
Sync
Reserved for future use
Connector
BNC female
Description
Specifications
Supplemental Information
Trigger Inputs
(Trigger 1 In, Trigger 2 In)
Either trigger source may be selected
Connector
BNC female
Impedance
10 k
Ω (nominal)
Trigger Level Range
−5 to +5 V
1.5 V (TTL) factory preset