Other connector rear, 2nd if output specifications, Aux if out port, Second if out – Agilent Technologies N9010A User Manual
Page 109: Aux if out port second if out
Chapter 6
109
Option CR3 - Connector Rear, 2nd IF Output
Other Connector Rear, 2nd IF Output Specifications
Other Connector Rear, 2nd IF Output Specifications
Aux IF Out Port
Second IF Out
Description
Specifications
Supplemental Information
Connector
SMA female
Shared with other options
Impedance
50
Ω (nominal)
Description
Specifications Supplemental Information
Second IF Out
Output Center Frequency
SA Mode
322.5 MHz
I/Q Analyzer Mode
IF Path
≤ 25 ΜΗz
322.5 MHz
IF Path
40 ΜΗz
250 MHz
Conversion Gain at 2nd IF
output center frequency
–1 to +4 dB (nominal) plus RF frequency response
a
a. “Conversion Gain” is defined from RF input to IF Output with 0 dB mechanical attenuation and the
electronic attenuator off. The nominal performance applies in zero span.
Bandwidth
Low band
Up to 140 MHz (nominal)
b
b. The passband width at –3 dB nominally extends from IF frequencies of 230 to 370 MHz. When using
IF paths with center frrequencies of 250 MHz or 322.5 MHz, the passband will therefore be asymmet-
ric.
High band
With preselector
Depends on RF center frequency
c
c. The YIG-tuned preselector bandwidth nominally varies from 55 MHz for a center frequencies of
3.6 GHz through 57 MHz at 15 GHz to 75 MHz at 26.5 GHz. (Refer to page 23 for details.) The prese-
lector effect will dominate the passband width.
Preselector bypassed
(Option MPB)
Up to 410 MHz nominal
d
d. The passband width at –6 dB nominally extends from 100 to 510 MHz. Thus, the maximum width is
not centered around the IF output center frequency.
Residual Output Signals