Electronic attenuator switching uncertainty, Electronic – Agilent Technologies N9010A User Manual
Page 120
120
Chapter 8
Option EA3 - Electronic Attenuator, 3.6 GHz
Other Electronic Attenuator Specifications
Description
Specifications
Supplemental Information
Electronic Attenuator Switching
Uncertainty
(Error relative to reference
condition: 50 MHz, 10 dB
mechanical attenuation, 10 dB
electronic attenuation)
Attenuation = 0 to 24 dB
9 kHz to 3.6 GHz
See note
a
a. The specification is
±0.14 dB. Note that this small relative uncertainty does not apply in estimating
absolute amplitude accuracy. It is included within the absolute amplitude accuracy for measurements
done with the electronic attenuator. (Measurements made without the electronic attenuator are treated
differently; the absolute amplitude accuracy specification for these measurements does not include
attenuator switching uncertainty.)