Chapter 1

31

Agilent EXA Signal Analyzer

Frequency and Time

c. Resolution Bandwidth Accuracy can be observed at slower sweep times than auto-coupled conditions.

Normal sweep rates cause the shape of the RBW filter displayed on the analyzer screen to widen by
nominally 6

%. This widening declines to 0.6% nominal when the Swp Time Rules key is set to Accu-

racy instead of Normal. The true bandwidth, which determines the response to impulsive signals and
noise-like signals, is not affected by the sweep rate.

Description

Specification

Supplemental information

Analysis Bandwidth

a

Standard

10 MHz

With Option B25

b

25 MHz

With Option B40

40 MHz

a. Analysis bandwidth is the instantaneous bandwidth available about a center frequency over which the

input signal can be digitized for further analysis or processing in the time, frequency, or modulation
domain.

b. Option B25 is standard for instruments ordered after May 1, 2011.

Description

Specifications

Supplemental Information

Preselector Bandwidth

Relevant to many options, such as B25 Wide IF
Bandwidth, in Bands 1 and higher. Nominal.

Mean Bandwidth at CF

a

Freq option

526

Freq option > 526

5 GHz

58 MHz

46 MHz

10 GHz

57 MHz

52 MHz

15 GHz

59 MHz

53 MHz

20 GHz

64 MHz

55 MHz

25 GHz

74 MHz

56 MHz

35 GHz

62 MHz

44 GHz

70 MHz

Standard Deviation

9%

7%

–3 dB Bandwidth

–7.5% relative to –4 dB bandwidth, nominal

a. The preselector can have a significant passband ripple. To avoid ambiguous results, the –4 dB band-

width is characterized.