beautypg.com

Adjacent channel power – Agilent Technologies N9010A User Manual

Page 268

background image

268

Chapter 30

LTE Measurement Application

Measurements

Description

Specifications

Supplemental Information

Adjacent Channel Power

Single Carrier

Minimum power at RF
input

−36 dBm (nominal)

Accuracy

Channel Bandwidth

Radio

Offset

5 MHz

10 MHz

20 MHz

ACPR Range for Specification

MS

Adjacent

a

±0.18 dB

±0.31 dB

±0.44 dB

−33 to −27 dBc with opt ML

b

BTS

Adjacent

c

±1.23 dB

±1.79. dB

±2.40 dB

−48 to −42 dBc with opt ML

d

BTS

Alternate

c

±0.29 dB

±0.46 dB

±0.87 dB

−48 to −42 dBc with opt ML

e

Dynamic Range E-UTRA

Test conditions

f

Offset

Channel
BW

Dynamic
Range
(nominal)

Optimum
Mixer Level
(nominal)

Adjacent

5 MHz

70.0 dB

−16.5 dBm

Adjacent

10 MHz

69.3 dB

−16.5 dBm

Adjacent

20 MHz

68.4

dB

−16.3dBm

Alternate

5 MHz

75.8 dB

−16.6 dBm

Alternate

10 MHz

73.2 dB

−16.4 dBm

Alternate

20 MHz

70.3 dB

−16.3 dBm

Dynamic Range UTRA

Test conditions

f

Offset

Channel
BW

Dynamic
Range
(nominal)

Optimum
Mixer Level
(nominal)

2.5 MHz

5 MHz

70.5 dB

−16.6 dBm

2.5 MHz

10 MHz

70.5 dB

−16.4 dBm

2.5 MHz

20 MHz

71.4 dB

−16.3 dBm

7.5 MHz

5 MHz

76.

5

dB

−16.6 dBm

7.5 MHz

10 MHz

76.5 dB

−16.4 dBm

7.5 MHz

20 MHz

75.7 dB

−16.3 dBm

a. Measurement bandwidths for mobile stations are 4.5, 9.0 and 18.0 MHz for channel bandwidths of 5,

10 and 20 MHz respectively.

b. The optimum mixer levels (ML) are

−22, −23 and −19 dBm for channel bandwidths of 5, 10 and 20

MHz respectively.

c. Measurement bandwidths for base transceiver stations are 4.515, 9.015 and 18.015 MHz for channel

bandwidths of 5, 10 and 20 MHz respectively.

d. The optimum mixer levels (ML) are

−18, −18 and −15 dBm for channel bandwidths of 5, 10 and 20

MHz respectively.

e. The optimum mixer level (ML) is

−8 dBm.

f. E-TM1.1 and E-TM1.2 used for test. Noise Correction set to On.