Chapter 1
Agilent EXA Signal Analyzer
Dynamic Range
Third Order Intermodulation
Description
Specifications
Supplemental Information
Third Order
Intermodulation
(Tone separation > 5 times IF
Prefilter Bandwidth
a
Verification conditions
b
)
Refer to the footnote for
.
Option 532, or 544 (mmW)
Option 503, 507, 513, or 526 (RF/
μW)
20 to 30
°C
Intercept
c
Intercept
(typical)
10 to 100 MHz
x
+12 dBm
+17 dBm
100 to 400 MHz
x
+10 dBm
+14 dBm
400 MHz to 1.7 GHz
x
+11 dBm
+15 dBm
1.7 to 3.6 GHz
x
+13 dBm
+17 dBm
100 MHz to 3.95 GHz
x
+15 dBm
+19 dBm
3.6 to 5.1 GHz
x
+11 dBm
+17 dBm
5.1 to 7 GHz
x
+13 dBm
+17 dBm
3.95 to 8.4 GHz
x
+15 dBm
+18 dBm
7 to 13.6 GHz
x
+11 dBm
+15 dBm
8.3 to 13.6 GHz
x
+15 dBm
+18 dBm
13.6 to 26.5 GHz
x
+9 dBm
+14 dBm
13.5 to 17.1 GHz
x
+11 dBm
+17 dBm
17.0 to 26.5 GHz
x
+10 dBm
+17 dBm (nominal)
26.5 to 44 GHz
x
+13 dBm (nominal)
Full temperature range
10 to 100 MHz
x
+10 dBm
100 to 400 MHz
x
+9 dBm
400 MHz to 1.7 GHz
x
+10 dBm
1.7 to 3.6 GHz
x
+12 dBm
100 MHz to 3.95 GHz
x
+13 dBm
3.6 to 5.1 GHz
x
+10 dBm
5.1 to 7 GHz
x
+12 dBm
3.95 to 8.4 GHz
x
+13 dBm
7 to 13.6 GHz
x
+10 dBm
8.3 to 13.6 GHz
x
+13 dBm