beautypg.com

Third order intermodulation – Agilent Technologies N9010A User Manual

Page 50

background image

50

Chapter 1

Agilent EXA Signal Analyzer

Dynamic Range

Third Order Intermodulation

Description

Specifications

Supplemental Information

Third Order

Intermodulation

(Tone separation > 5 times IF

Prefilter Bandwidth

a

Verification conditions

b

)

Refer to the footnote for

Band Overlaps on page 21

.

Option 532, or 544 (mmW)

Option 503, 507, 513, or 526 (RF/

μW)

20 to 30

°C

Intercept

c

Intercept
(typical)

10 to 100 MHz

x

+12 dBm

+17 dBm

100 to 400 MHz

x

+10 dBm

+14 dBm

400 MHz to 1.7 GHz

x

+11 dBm

+15 dBm

1.7 to 3.6 GHz

x

+13 dBm

+17 dBm

100 MHz to 3.95 GHz

x

+15 dBm

+19 dBm

3.6 to 5.1 GHz

x

+11 dBm

+17 dBm

5.1 to 7 GHz

x

+13 dBm

+17 dBm

3.95 to 8.4 GHz

x

+15 dBm

+18 dBm

7 to 13.6 GHz

x

+11 dBm

+15 dBm

8.3 to 13.6 GHz

x

+15 dBm

+18 dBm

13.6 to 26.5 GHz

x

+9 dBm

+14 dBm

13.5 to 17.1 GHz

x

+11 dBm

+17 dBm

17.0 to 26.5 GHz

x

+10 dBm

+17 dBm (nominal)

26.5 to 44 GHz

x

+13 dBm (nominal)

Full temperature range

10 to 100 MHz

x

+10 dBm

100 to 400 MHz

x

+9 dBm

400 MHz to 1.7 GHz

x

+10 dBm

1.7 to 3.6 GHz

x

+12 dBm

100 MHz to 3.95 GHz

x

+13 dBm

3.6 to 5.1 GHz

x

+10 dBm

5.1 to 7 GHz

x

+12 dBm

3.95 to 8.4 GHz

x

+13 dBm

7 to 13.6 GHz

x

+10 dBm

8.3 to 13.6 GHz

x

+13 dBm