Agilent Technologies N9010A User Manual
Page 51
Chapter 1
51
Agilent EXA Signal Analyzer
Dynamic Range
13.6 to 26.5 GHz
x
+7 dBm
13.5 to 17.1 GHz
x
+9 dBm
17.0 to 26.5 GHz
x
+8 dBm
a. See the IF Prefilter Bandwidth table in the Gain Compression specifications on
. When the tone
separation condition is met, the effect on TOI of the setting of IF Gain is negligible. TOI is verified
with IF Gain set to its best case condition, which is IF Gain = Low.
b. TOI is verified with two tones, each at
−16 dBm at the mixer, spaced by 100 kHz.
c. Intercept = TOI = third order intercept. The TOI is given by the mixer tone level (in dBm) minus (dis-
tortion/2) where distortion is the relative level of the distortion tones in dBc.
Nominal Dynamic Range vs. Offset Frequency vs. RBW for Freq Option
≤ 526 [Plot]
Description
Specifications
Supplemental Information