Displayed average noise level, Displayed average noise level (danl) – Agilent Technologies N9010A User Manual
Page 46
46
Chapter 1
Agilent EXA Signal Analyzer
Dynamic Range
Displayed Average Noise Level
Description
Specifications
Supplemental
Information
Displayed Average Noise Level
(DANL)
a
Input terminated
Sample or Average detector
Averaging type = Log
0 dB input attenuation
IF Gain = High
1 Hz Resolution Bandwidth
Refer to the footnote
for
.
mmW without Option B40, DP2, or MPB
mmW with Option B40, DP2, or MPB
RF/
μW(Option 503, 507, 513, or 526)
20 to 30
°C
Full range
Typical
10 Hz
x
x
x
–90 dBm (nominal)
20 Hz
x
x
x
–100 dBm (nominal)
100 Hz
x
x
x
–110 dBm (nominal)
1 kHz
x
x
x
–120 dBm (nominal)
9 kHz to 1 MHz
x
–125 dBm (nominal)
9 kHz to 1 MHz
x
x
−130 dBm
1 to 10 MHz
b
x
−147 dBm
−145 dBm
−149 dBm
1 MHz to 1.2 GHz
x
x
−152 dBm
−151 dBm
−155 dBm
10 MHz to 2.1 GHz
x
−148 dBm
−146 dBm
−150 dBm
1.2 to 2.1 GHz
x
x
−151 dBm
−150 dBm
−154 dBm
2.1 to 3.6 GHz
x
−147 dBm
−145 dBm
−149 dBm
2.1 to 3.6 GHz
x
x
−149 dBm
−148 dBm
−152 dBm
3.6 to 7 GHz
x
−147 dBm
−145 dBm
−149 dBm
3.5 to 4.2 GHz
x
−142 dBm
−140 dBm
−146 dBm
3.5 to 4.2 GHz
x
−144 dBm
−142 dBm
−147 dBm
4.2 to 8.4 GHz
x
−143 dBm
−141 dBm
−148 dBm
4.2 to 8.4 GHz
x
−145 dBm
−143 dBm
−150 dBm
7 to 13.6 GHz
x
−143 dBm
−141 dBm
−147 dBm
8.3 to 13.6 GHz
x
−145 dBm
−143 dBm
−148 dBm
8.3 to 13.6 GHz
x
−147 dBm
−145 dBm
−150 dBm
13.5 to 22 GHz
x
−137 dBm
−134 dBm
−142 dBm
13.5 to 20 GHz
x
−142 dBm
−140 dBm
−146 dBm
13.5 to 20 GHz
x
−145 dBm
−143 dBm
−148 dBm