beautypg.com

Agilent Technologies N9010A User Manual

Page 324

background image

324

Chapter 35

WLAN Measurement Application

List Sequence Measurements

Description

Specifications

Supplemental Information

64QAM EVMRF Input
Level = –10 dBm,
Attenuation = 10 dB,

20 to 30°C

Radio standards are:

802.11a/g/j/p (OFDM),

802.11g (DSSS-OFDM),

802.11n (20 MHz) or

802.11ac (20 MHz)

Center Frequency in 5.0 GHz Band

Code Rate: 3/4

EQ Training: Channel Est Seq Only

Track Phase On

Track Amp Off

Track Timing Off

EVM

Floor

abc

–47.0 dB (0.45%) (nominal)

Accuracy

(EVM Range:0 to 8.0%)

±0.30% (nominal)

Frequency Error

Range

±100 kHz (nominal)

Accuracy

±10 Hz + tfa

d

(nominal)

a. In these specifications, those values with dB units are the specifications, while those with % units, in

parentheses, are conversions from the dB units to % for reader convenience.

b. The EVM Floor specification applies when Phase Noise Optimization is set to Wide-offset (>30 kHz)
c. The EVM Floor specificaton applies when

μW Path Control is set to μW Preselector Bypass.

d. tfa = transmitter frequency

×

frequency reference accuracy.