Adjacent channel power (acp) – Agilent Technologies N9010A User Manual
Page 59
![background image](/manuals/35678/59/background.png)
Chapter 1
59
Agilent EXA Signal Analyzer
Power Suite Measurements
Description
Specifications
Supplemental Information
Adjacent Channel Power (ACP)
Case: Radio Std = None
Accuracy of ACP Ratio (dBc)
Display Scale Fidelity
a
Accuracy of ACP Absolute Power
(dBm or dBm/Hz)
Absolute Amplitude Accuracy
b
+
Power Bandwidth Accuracy
cd
Accuracy of Carrier Power (dBm), or
Carrier Power PSD (dBm/Hz)
Absolute Amplitude Accuracy
+
Power Bandwidth Accuracy
Passband Width
e
−3 dB
Case: Radio Std = 3GPP W-CDMA
(ACPR; ACLR)
f
Minimum power at RF Input
−36 dBm (nominal)
ACPR Accuracy
g
RRC weighted, 3.84 MHz noise
bandwidth, method
≠ RBW
Radio
Offset Freq
MS (UE)
5 MHz
±0.22 dB
At ACPR range of
−30 to −36 dBc with
optimum mixer level
h
MS (UE)
10 MHz
±0.34 dB
At ACPR range of
−40 to −46 dBc with
optimum mixer level
i
BTS
5 MHz
±1.07 dB
At ACPR range of
−42 to −48 dBc with
optimum mixer level
j
BTS
10 MHz
±1.00 dB
At ACPR range of
−47 to −53 dBc with
optimum mixer level
BTS
5 MHz
±0.44 dB
At
−48 dBc non-coherent ACPR
k
Dynamic Range
RRC weighted, 3.84 MHz noise
bandwidth
Noise
Correction
Offset
Freq
Method
ACLR (typical)
l
Optimum ML
m
(Nominal)
Off
5 MHz Filtered IBW
−68 dB
−8 dBm
Off
5 MHz Fast
−67 dB
−9 dBm
Off
10 MHz Filtered IBW
−74 dB
−2 dBm
On
5 MHz Filtered IBW
−73 dB
−8 dBm
On
10 MHz Filtered IBW
−76 dB
−2 dBm
RRC Weighting Accuracy
n
White noise in Adjacent Channel
TOI-induced spectrum
rms CW error
0.00 dB nominal
0.001 dB nominal
0.012 dB nominal
a. The effect of scale fidelity on the ratio of two powers is called the relative scale fidelity. The scale
fidelity specified in the Amplitude section is an absolute scale fidelity with –35 dBm at the input mixer
as the reference point. The relative scale fidelity is nominally only 0.01 dB larger than the absolute
scale fidelity.
b. See Amplitude Accuracy and Range section.
c. See Frequency and Time section.