beautypg.com

Edr modulation accuracy – Agilent Technologies N9010A User Manual

Page 201

background image

Chapter 21

201

Bluetooth Measurement Application

Enhanced Data Rate (EDR) Measurements

Description

Specifications

Supplemental Information

EDR Modulation Accuracy

This measurement is a Transmit
Analysis measurement and is in
conformance with Bluetooth RF
test specification 2.1.E.0.5.1.13

Packet Type

2-DH1, 2-DH3, 2-DH5, 3-DH1,
3-DH3, 3-DH5

Payload

PRBS9, BS00, BSFF, BS55

Synchronization

DPSK synchronization sequence

Trigger

External, RF Burst, Periodic
Timer, Free Run, Video

Supported measurements

rms DEVM

peak DEVM, 99% DEVM

RF input level range

a

+30 dBm to –70 dBm

RMS DEVM

Range

0 to 12%

Floor

1.5%

Accuracy

b

1.2%

a. When the input signal level is lower than –40 dBm, the analyzer’s preamp should be turned on and the

attenuator set to 0 dB.

b. The accuracy specification applies when the EVM to be measured is well above the measurement

floor. When the EVM does not greatly exceed the floor, the errors due to the floor add to the accuracy
errors. The errors due to the floor are noise-like and add incoherently with the UUT EVM. The errors
depend on the EVM of the UUT and the floor as follows:

error = sqrt(EVMUUT

2

+ EVMsa

2

) – EVMUUT, where EVMUUT is the EVM of the UUT in percent,

and EVMsa is the EVM floor of the analyzer in percent