Chapter 1
69
Agilent EXA Signal Analyzer
General
Description
Specification
Supplemental Information
Power Requirements
Low Range
Voltage
100 to 120 V
Frequency
Serial Prefix < MY4801,
SG4801, or US4801
50 or 60 Hz
Serial Prefix
≥ MY4801,
SG4801, or US4801
50, 60 or 400 Hz
High Range
Voltage
220 to 240 V
Frequency
50 or 60 Hz
Power Consumption, On
350 W
Maximum
Power Consumption, Standby
20 W
Standby power is not supplied to
frequency reference oscillator.
Typical instrument configuration
Power (nominal)
Base 3.6 GHz instrument (N9010A-503)
176 W
Base 8.4 GHz instrument (N9010A-508)
179 W
Base 13 GHz instrument (N9010A-513)
183 W
Base 26.5 GHz instrument (N9010A-526)
194 W
Base 32/44 GHz instrument (N9010A-532/544)
225 W
Adding Option B40, MPB, or DP2 to base
instrument
+45 W
Description
Supplemental Information
Measurement Speed
a
a. Sweep Points = 101.
Nominal
Standard
w/ Option PC4
Local measurement and display update rate
bc
11 ms (90/s)
4 ms (250/s)
Remote measurement and LAN transfer rate
6 ms (167/s)
5 ms (200/s)
Marker Peak Search
5 ms
1.5 ms
Center Frequency Tune and Transfer (RF)
22 ms
20 ms
Center Frequency Tune and Transfer (µW)
49 ms
47 ms
Measurement/Mode Switching
75 ms
39 ms
W-CDMA ACLR measurement time
See
Measurement Time vs. Span
See