Chapter 16

Analog Demodulation Measurement Application

FM Stereo/Radio Data System (RDS) Measurements

FM Stereo/Radio Data System (RDS) Measurements

1

1. The N9063A-3FP option requires that the instrument also has the N9063A-2FP option installed and licensed.

Description

Specifications

Supplemental Information

FM Stereo Modulation
Analysis Measurements

MXP view

RF Spectrum, AF Spectrum, Demod
Waveform, FM Deviation (Hz) (Peak
+, Peak–, (Pk-Pk)/2, RMS), Carrier
Power (dBm), Carrier Frequency Error
(Hz), SINAD (dB), Distortion (% or
dB)

MPX consists of FM signal
multiplexing with the mono signal
(L+R), stereo signal (L–R), pilot
signal (at 19 kHz) and optional
RDS signal (at 57 kHz).

SINAD MPX BW, default 53
kHz, range from 1 kHz to 58
kHz.

Reference Deviation, default 75
kHz, range from 15 kHz to 150
kHz.

Mono (L+R) / Stereo
(L–R) view

Demod Waveform, AF Spectrum,
Carrier Power (dBm), Carrier
Frequency Error (Hz), Modulation
Rate

Mono Signal is Left + Right

Stereo Signal is Left – Right

Left / Right view

Demod Waveform, AF Spectrum,
Carrier Power (dBm), Carrier
Frequency Error (Hz), Modulation
Rate, SINAD (dB), Distortion (% or
dB), THD (% or dB)

Post-demod settings:

Highpass filter: 20, 50, or 300
Hz

Lowpass filter: 300 Hz, 3, 15,
80, or 300 kHz

Bandpass filter: A-Weighted,
CCITT

De-Emphasis: 25, 50, 75 and

750

μs

RDS / RBDS Decoding
Results view

BLER basic tuning and switching
information, radio text, program item
number and slow labeling codes, clock
time and date

BLER Block Count default 1E+8,