Using the device ports dialog, 3 port configuration for infusion – Teledyne LeCroy Sierra M122 User Manual User Manual
Page 229

Sierra M122 SAS/SATA Protocol Analyzer User Manual
227
Port Configuration for InFusion
Teledyne LeCroy
Using the first row of icons, you can Run/Stop All Ports or Run/Stop individual ports.
The columns display the Port, Function/Configuration, Library, and Scenario.
You can Float, Dock, Auto‐Hide, or Hide the window.
Note:
A port row is grayed‐out when that port has not been configured to be a Jammer in the
Port Configuration dialog (see
“Port Configuration for InFusion” on page 227
Using the Device Ports Dialog
After you have finished Port Configuration (see
“Port Configuration for InFusion” on
), you use the Device Ports dialog to assign specific scenarios to ports, so that
different scenarios can run on different ports.
To assign a scenario to a port, drag and drop the scenario from any library window to the
port. The Device Ports dialog then displays the Library and Scenario on the row for that
Port/Configuration.
Figure 4.5: Device Ports Dialog
Alternatively, assign the scenario using the Library and Scenario drop‐down lists.
Figure 4.6: Scenario Drop-down List
After you have assigned scenarios to ports, in the first row of icons, use the first green
arrow icon to Run/Stop All Ports, or use the numbered green arrows to Run/Stop an
individual port.
Note:
A port row is grayed‐out when that port is running a scenario.
4.3
Port Configuration for InFusion
The InFusion (Jammer) port configurations must match the Analyzer port configurations
for the infusion‐analyzer to work.
Select Configuration > Port Configuration to display the Set Port Configuration dialog.