Teledyne LeCroy

TxRxVout & Preemphasis

198

 

 

Sierra M122 SAS/SATA Protocol Analyzer User Manual

To select TX Vout, select Setup > TxRx Vout & Preemphasis to display the 
Rx/Tx Settings dialog:

Figure 3.128: Choose Port for TxRx Vout

Port displays ports to select from. 

Copy Selected Port Settings to All Ports implements selected port's setting into all other 
port settings.

PMA Analog Control Settings allows you to select values for RX equalization DC and RX 
equalization (Write Only)

Advanced displays the Advanced Probe Setting dialog (see 

“Advanced Probe Setting” on 

page 199

).

References displays the References Probe Setting dialog (see 

“References Probe Setting” 

on page 199

).

Overwrite Tx Settings check this box to overwrite the Tx settings.

Disable Rx Training check this box to use the manually entered settings.

Port Status displays number of errors per second.

Apply applies the selected settings.

Start Reading Port Status implements reading of number of errors displayed in Port 
Status. 

Restore Factory Settings restores default values.

Save saves the new values as a *.sng file.

Load loads back the saved *.sng file.

Auto Calibration displays the Auto Probe Calibration dialog (see 

“Auto Probe Calibration” 

on page 202

).