TxRxVout & Preemphasis

200

 

 

Sierra M122 SAS/SATA Protocol Analyzer User Manual

Figure 3.130: References Probe Setting Dialog

3.15.3

Manual Calibration of Tx Path from Jammer

To manually calibrate perform the following steps:

1. Select Setup > TxRx Vout & Preemphasis to display the 

Rx/Tx Settings dialog (see 

Figure 3.131 on page 200

).

2. Press the Read button to read the current values for all ports (values achieved 

through automatic Training when the link has come up), then press the Advanced 
button to see and tweak the values.

3. The typical set of values for the Initiator are shown in 

Figure 3.132 on page 201

. If 

you see these values then no changes are required.

Figure 3.131: Rx/Tx Settings Dialog