Operating Manual for

XDC Series Power Sup

ply

SCPI Command Re

fe

rence

SCPI C

omman

d Summary

Table B.15Legacy Commands

Notes:

All legacy commands that change a value conform to the same rules as SCPI. Commands will cause an error if the unit’s remote
source, remote state, current share mode status and calibration status is incorrect.

Legacy commands do not have multichannel capabilities.

Query commands return a string containing the command itself.

The “CLR” command does not clear any legacy fault registers.

The “RST” command only clears the protections so that the voltage and setpoints can take effect. It does not change any setpoint
values. When a protection is tripped the unit’s output may be turned off (configurable).

Function

Legacy Command

Description

Query

Reset

CLR

Equivalent to *RST and SYSTem:RESet commands
Performs a device reset. Set the power supply to a known state that is independent of
the use history of the device

N/A

Query System Error

ERR?

Equivalent to SYSTem:ERRor? command except that the return string contains the
command
Returns the next error in the instrument’s error queue

N/A

Identification Query

ID?

Query identification string. (Model ID and Version)

N/A

Read Output Current

IOUT?

Equivalent to MEASure:CURRent? command
Read output current

N/A

Set High Current Limit

IMAX

Equivalent to SOURce:CURRent:LIMit:HIGH command.
Set upper limit of current setpoint range (soft limits)

Yes

Set (Immediate) Current
Setpoint

ISET

Equivalent to SOURce:CURRent command.
Change current setpoint

Yes

Set Output State

OUT

Equivalent to OUTP command.
Enable/disable the power supply output

Yes

Set Over Voltage
Protection Level

OVSET

Equivalent to SOURce:VOLTage:PROTection:OVER:LEVel command.
Set the over voltage protection level

Yes

ROM Query

ROM?

Queries the main firmware version

N/A

Clear Output Protection

RST

Equivalent to OUTPut:PROTection:CLEar command.
Clears the protection mechanism

N/A

Set High Voltage Limit

VMAX

Equivalent to SOURce:VOLTage:LIMit:HIGH command.
Set upper limit of voltage setpoint range (soft limits)

Yes

Read Output Voltage

VOUT?

Equivalent to MEASure:VOLTage? command.
Read output voltage

N/A

Set (Immediate) Voltage
Setpoint

VSET

Equivalent to SOURce:VOLTage command.
Change voltage setpoint

Yes