Change remote/local control of power supply, Enable output, Program v,i,p – Xantrex Technology XDC User Manual
Page 130: Scpi, Enable output program v,i,p

Remote Operation
SCPI Commands for Digital Interfaces
128
Operating Manual for XDC Series Power Supply
Change
Remote/Local
Control of
Power Supply
A SCPI command is provided for use with the RS-232 and multichannel interfaces
to change the remote/local mode. (GPIB will use IEEE 4888-1 functions to change
modes.)
SYST:REM:STAT {LOC|REM|RWL}
Where:
•
LOC: go to local mode operation
•
REM: go to remote mode operation
•
RWL: remote with local lockout. Go to remote mode operation with local mode
locked out. With RWL set, the user cannot return to local mode via the front
panel.
To query to remote mode:
SYST:REM:STAT?
Enable
Output
To enable or disable the output:
OUTP {ON|OFF}
To query the state of the output enable:
OUTP?
Program V,I,P
SCPI
To change setpoints:
SOUR:VOLT
SOUR:CURR
SOUR:POW
To check setpoints:
SOUR:VOLT?
SOUR:CURR?
SOUR:POW?
To set a triggered setpoint:
SOUR:VOLT:TRIG
SOUR:CURR:TRIG
SOUR:POW:TRIG
See “Triggering Commands” on page 136 for more information.