Change remote/local control of power supply, Enable output, Program v,i,p – Xantrex Technology XDC User Manual

Page 130: Scpi, Enable output program v,i,p

background image

Remote Operation
SCPI Commands for Digital Interfaces

128

Operating Manual for XDC Series Power Supply

Change

Remote/Local

Control of

Power Supply

A SCPI command is provided for use with the RS-232 and multichannel interfaces
to change the remote/local mode. (GPIB will use IEEE 4888-1 functions to change
modes.)

SYST:REM:STAT {LOC|REM|RWL}

Where:

LOC: go to local mode operation

REM: go to remote mode operation

RWL: remote with local lockout. Go to remote mode operation with local mode

locked out. With RWL set, the user cannot return to local mode via the front

panel.

To query to remote mode:
SYST:REM:STAT?

Enable
Output

To enable or disable the output:
OUTP {ON|OFF}

To query the state of the output enable:
OUTP?

Program V,I,P

SCPI

To change setpoints:
SOUR:VOLT
SOUR:CURR
SOUR:POW

To check setpoints:
SOUR:VOLT?
SOUR:CURR?
SOUR:POW?

To set a triggered setpoint:
SOUR:VOLT:TRIG
SOUR:CURR:TRIG
SOUR:POW:TRIG

See “Triggering Commands” on page 136 for more information.