Calibration
Entering Calibration Mode

180

Operating Manual for XDC Series Power Supply

Front Panel To access calibration mode via the front panel

:

If your password code is incorrect, the prompt displays Incorrect code and

automatically returns to Code ####. Either try again or press EXIT to escape.

If your password code is correct, the prompt displays Output V Cal. This is the

first of 12 available options. The 12 sets of parameters that must be adjusted during
calibration are:

Output V Cal: Calibrate voltage output and readback.

Output I Cal: Calibrate current output and readback.

ANLG V PGM 5V: Calibrate 5V analog programming interface for setting

voltage output.

ANLG V RB 5V: Calibrate 5V analog programming interface for monitoring

voltage output.

ANLG I PGM 5V: Calibrate 5V analog programming interface for setting

current output.

ANLG I RB 5V: Calibrate 5V analog programming interface for monitoring

current output.

ANLG V PGM 10V: Calibrate 10V analog programming interface for setting

voltage output.

ANLG V RB 10V: Calibrate 10V analog programming interface for monitoring

voltage output.

ANLG I PGM 10V: Calibrate 10V analog programming interface for setting

current output.

ANLG I RB 10V: Calibrate 10V analog programming interface for monitoring

current output.

Factory Cal: Lets you restore the factory calibration constants.

Change Code: Lets you change the password code.

Step # Do This

You Will See

1

ERROR MSGS

2

Press 2 times.

CALIBRATION

3

Code #####

4

Enter the calibration security code.
The factory code is “0000”.

Code 0000

5

You are now in the

Calibration menu.

Output V Cal

MENU

ENTER

ENTER