Calibration
Front Panel Calibration Procedure

186

Operating Manual for XDC Series Power Supply

4. Enter voltage data

Enter the voltage at the current programming lines, read from the external DVM.

Press ENTER

5. Maximum calibration level

Set the input to the programming lines to approximately 4.5V (90% of full

scale).

6. Enter voltage data

Enter the voltage at the current programming lines, read from the external DVM.

Press ENTER.

7. Power supply calculates and stores calibration constants.
8. The menu will go to the 5V analog current readback calibration menu.

Analog Programming Interface Current Readback Calibration

1. Attach a DVM across the current readback lines, Pins B6 and B1(GND).
2. Enter 5V analog current readback calibration menu

Select ANLG I RB 5V from the calibration menu. Press ENTER.

You will be prompted to set up for analog current readback calibration. Press

ENTER when ready.

3. Minimum calibration level

The power supply will automatically set the current readback lines to

approximately 10% of full scale.

4. Enter voltage data

Enter the voltage read from the external DVM. Press ENTER

5. Maximum calibration level

The power supply will automatically set the current readback lines to

approximately 90% of full scale.

6. Enter voltage data

Enter the voltage at the current readback lines, read from the external DVM.

Press ENTER.

7. Power supply calculates and stores calibration constants.
8. The menu will go to the 10V analog voltage programming calibration menu.