beautypg.com

Table 4.8, Features affected by reset (*rst) command – Xantrex Technology XDC User Manual

Page 128

background image

Remote Operation
SCPI Commands for Digital Interfaces

126

Operating Manual for XDC Series Power Supply

Table 4.8 Features Affected by Reset (*RST) Command

Feature

Reset State (*RST)

Voltage setpoint

0.0V

Current setpnoint

0.0A

Power setpoint

103% of power rating

Low voltage setpoint limit 0.0V
High voltage setpoint limit 103% of voltage rating
Low current setpoint limit 0.0A
High current setpoint limit 103% of current rating
Low power setpoint

3% of power rating

High power setpoint

103% of power rating

Over voltage protection

Disabled (0.0V)

Under voltage protection

Disabled (0.0V) and not shutdown when tripped

Over current protection

Disabled (0.0A) and not shutdown when tripped

Under current protection

Disabled (0.0A) and not shutdown when tripped

Over power protection

Disabled (0.0W) and not shutdown when tripped

Under power protection

Disabled (0.0W) and not shutdown when tripped

Fold shutdown protection None and delay 0.5s
AC off shutdown recovery Autorecover
Over temperature
shutdown recovery

Latched

Triggered voltage
setpoint

Disabled (Default)

Triggered current setpoint Disabled (Default)
Triggered power setpoint Disabled (Default)
Autosequence mode

Exit

Front panel display config Show V & I
Knob lockout

None

Calibration mode

False

Output

Disabled