Exit calibration mode, Restore factory calibration, Note – Xantrex Technology XDC User Manual
Page 194: Exit calibration mode restore factory calibration
Calibration
Exit calibration mode
192
Operating Manual for XDC Series Power Supply
Exit calibration mode
When you have completed calibration, hit the EXIT key.
The SCPI command is:
CAL:STAT OFF, "0000"
Restore Factory Calibration
To restore the unit to the calibration constants set at the factory:
The SCPI command is
CALibration:RESTore
Step # Do This
You Will See
1
Output V Cal
2
or
Scroll to select Factory Cal.
Factory Cal
3
Select Y to Restore the factory
calibration. Selecting N will exit.
Restore? Y
Note This procedure should not be used in place of regular calibration, but may be useful to
restore the unit to an operational state in case of failure.