Calibration
Remote Interface Calibration Procedure

190

Operating Manual for XDC Series Power Supply

3. Enter voltage data

Enter the voltage at the current programming lines, read from the external DVM.

CAL:ANAL:5V:PROG:CURR:DATA

4. Maximum calibration level

Set the input to the programming lines to approximately 4.5V (90% of full

scale).

Set the power supply to receive 5V analog current programming calibration data

by sending the command:

CAL:ANAL:5V:PROG:CURR:LEV MAX

5. Enter voltage data

Enter the voltage at the current programming lines, read from the external DVM.

CAL:ANAL:5V:PROG:CURR:DATA

6. Power supply calculates and stores calibration constants.

Analog Programming Interface Current Readback Calibration

1. Attach a DVM across the current readback lines, Pins B6 and B1(GND).
2. Minimum calibration level

Set the current readback lines to approximately 10% of full scale by sending the

command:

CAL:ANAL:5V:READ:CURR:LEV MIN

3. Enter voltage data

Enter the voltage across the current readback lines, read from the external DVM.

CAL:ANAL:5V:READ:CURR:DATA

4. Maximum calibration level

Set the current readback lines to approximately 90% of full scale by sending the

command:

CAL:ANAL:5V:READ:CURR:LEV MAX

5. Enter voltage data

Enter the voltage across the current readback lines, read from the external DVM.

CAL:ANAL:5V:READ:CURR:DATA

6. Power supply calculates and stores calibration constants.