Calibration

Remote Interface Calibration Procedure

Release 3.0

189

5. Enter voltage data

Enter the voltage at the voltage programming lines, read from the external DVM.

CAL:ANAL:5V:PROG:VOLT:DATA

6. Power supply calculates and stores calibration constants.

Analog Programming Interface Voltage Readback Calibration

1. Attach a DVM across the voltage readback lines, Pins B5 and B1(GND).
2. Minimum calibration level

Set the voltage readback lines to approximately 10% of full scale by sending the

command:

CAL:ANAL:5V:READ:VOLT:LEV MIN

3. Enter voltage data

Enter the voltage across the voltage readback lines, read from the external DVM.

CAL:ANAL:5V:READ:VOLT:DATA

4. Maximum calibration level

Set the voltage readback lines to approximately 90% of full scale by sending the

command:

CAL:ANAL:5V:READ:VOLT:LEV MAX

5. Enter voltage data

Enter the voltage across the voltage readback lines, read from the external DVM.

CAL:ANAL:5V:READ:VOLT:DATA

6. Power supply calculates and stores calibration constants.

Analog Programming Interface Current Programming Calibration

1. Connect the power source across the current programming lines, Pins B4 and

B1(GND). Attach a DVM across the current programming lines as well.

2. Minimum calibration level

Set the input to the programming lines to approximately 0.5V (10% of full

scale).

Set the power supply to receive 5V analog current programming calibration data

with the command:

CAL:ANAL:5V:PROG:CURR:LEV MIN