Enter voltage data, Minimum calibration level, Maximum calibration level – Xantrex Technology XDC User Manual
Page 191

Calibration
Remote Interface Calibration Procedure
Release 3.0
189
5. Enter voltage data
Enter the voltage at the voltage programming lines, read from the external DVM.
CAL:ANAL:5V:PROG:VOLT:DATA
6. Power supply calculates and stores calibration constants.
Analog Programming Interface Voltage Readback Calibration
1. Attach a DVM across the voltage readback lines, Pins B5 and B1(GND).
2. Minimum calibration level
Set the voltage readback lines to approximately 10% of full scale by sending the
command:
CAL:ANAL:5V:READ:VOLT:LEV MIN
3. Enter voltage data
Enter the voltage across the voltage readback lines, read from the external DVM.
CAL:ANAL:5V:READ:VOLT:DATA
4. Maximum calibration level
Set the voltage readback lines to approximately 90% of full scale by sending the
command:
CAL:ANAL:5V:READ:VOLT:LEV MAX
5. Enter voltage data
Enter the voltage across the voltage readback lines, read from the external DVM.
CAL:ANAL:5V:READ:VOLT:DATA
6. Power supply calculates and stores calibration constants.
Analog Programming Interface Current Programming Calibration
1. Connect the power source across the current programming lines, Pins B4 and
B1(GND). Attach a DVM across the current programming lines as well.
2. Minimum calibration level
Set the input to the programming lines to approximately 0.5V (10% of full
scale).
Set the power supply to receive 5V analog current programming calibration data
with the command:
CAL:ANAL:5V:PROG:CURR:LEV MIN