Table 4.15 QUEStionable Status Register
Bit
Bit Weight
Bit Name
Description
0
1
VOLTage Summary
Reflects a summary of the
VOLTage Sub-Register.
1
2
CURRent Summary
Reflects a summary of the
CURRent Sub-Register.
2
4
TIME Not
implemented
3
8
POWer Summary
Reflects a summary of the POWer
Sub-Register.
4
16
TEMPerature Summary
Reflects a summary of the
TEMPerature Sub-Register.
5 32
FREQuency
Summary
Not
implemented
6
64
PHASe Summary
Not implemented
7
128
MODulation Summary
Not implemented
8
256
CALibration
Indicates an error in the unit
calibration.
9
512
Not Used
Not implemented
10
1024
Not Used
Not implemented
11
2048
AC Off
Indicates an AC Supply failure.
12
4096
UNRegulated
Indicates that the output is not
regulated in either Constant
Voltage mode, Constant Current
mode or Constant Power mode.
Reflects the inverse of the
Operation Regulation Summary bit.
13
8192
INSTrument Summary
Not implemented
14
16384
Command Warning
Not implemented
15
32768
Not Used
Always zero