Remote Operation

SCPI Commands for Digital Interfaces

Release 3.0

131

Under-Current Protection

To set the Under-Current Protection level:
SOUR:CURR:PROT:UND
SOUR:CURR:PROT:UND:STAT

The first UCP protection command sets the protection level. The second command
lets you choose a warning alarm only [OFF] or shut down with an alarm [ON] if the
protection level is exceeded.

To check the Under-Current Protection level:
SOUR:CURR:PROT?

To check if the Under-Current Protection was tripped:
SOUR:CURR:PROT:TRIP?

Over-Power Protection

To set the Over-Power Protection level:
SOUR:POW:PROT
SOUR:POW:PROT:STAT

The first OPP protection command sets the protection level. The second command
lets you choose a warning alarm only [OFF] or shut down with an alarm [ON] if the
protection level is exceeded.

To check the Over-Power Protection level:
SOUR:POW:PROT?

To check if the Over-Power Protection was tripped:
SOUR:POW:PROT:TRIP?

Under-Power Protection

To set the Under-Power Protection level:
SOUR:POW:PROT:UND
SOUR:POW:PROT:UND:STAT

The first UPP protection command sets the protection level. The second command
lets you choose a warning alarm only [OFF] or shut down with an alarm [ON] if the
protection level is exceeded.

To check the Under-Power Protection level:
SOUR:POW:PROT?

To check if the Under-Power Protection was tripped:
SOUR:POW:PROT:TRIP?