G Program Selection Method Setup

The program selection method can be set to the front panel or the rear terminals.

Setting Procedure
1. Select the OPTIONS screen. Press the [OPTIONS] button.

2. Select the UTILITY screen. Press the [UTILITY] button.

3. Set the program selection method.

Press the [PROGRAM SELECT] button to select [PANEL] or [EXT].

Every time you press the button, the display toggles between [PANEL] and
[EXT].

Note 1: When the program selection method is set to [PANEL], program selection
can be performed using the front panel and the RS-232C input.

Note 2: When the program selection method is set to [EXT], program selection can
be performed using the terminal input only.

Note 3: When the panel lock function is used, program selection is performed as
shown below.

➮ For information on the panel lock function, see p. 112.

OPTIONS

UTILITY

LASER ON

OPTIONS

NOR

MAL

P. 1

DISPLAY

DATA

LIMIT

SETUP

PROGRAM

CALIB

OPTIONS

AREA

OUTPUT

INSPECT

UTILITY

Select Option to setup.

MODE

CHANGE

Normal

-> Pitch

LASER ON

UTILITY

NOR

MAL

P. 1

AREA CHECK

INITIALIZE

ON

PANEL

mm

ONE-SCAN MODE

PROGRAM SELECT

DISPLAY UNIT

ESC

Program selection

Panel

Terminal

Front panel

Enabled

Disabled

Terminal input

Disabled

Enabled

RS-232C input

Enabled

Disabled

Program selection

Panel

Terminal

Front panel

Disabled

Disabled

Terminal input

Disabled

Enabled

RS-232C input

Enabled

Disabled