17 program selection method setup, Gg program selection method setup, Chapter 6 normal mode – KEYENCE LS-5000 User Manual
Page 70

Chapter 6 Normal Mode
64
6
G
G
G
G
G Program Selection Method Setup
The program selection method can be set to the front panel or the rear terminals.
■ Setting Procedure
1. Select the OPTIONS screen. Press the [OPTIONS] button.
2. Select the UTILITY screen. Press the [UTILITY] button.
3. Set the program selection method.
Press the [PROGRAM SELECT] button to select [PANEL] or [EXT].
Every time you press the button, the display toggles between [PANEL] and
[EXT].
Note 1: When the program selection method is set to [PANEL], program selection
can be performed using the front panel and the RS-232C input.
Note 2: When the program selection method is set to [EXT], program selection can
be performed using the terminal input only.
Note 3: When the panel lock function is used, program selection is performed as
shown below.
➮ For information on the panel lock function, see p. 112.
OPTIONS
UTILITY
LASER ON
OPTIONS
NOR
MAL
P. 1
DISPLAY
DATA
LIMIT
SETUP
PROGRAM
CALIB
OPTIONS
AREA
OUTPUT
INSPECT
UTILITY
Select Option to setup.
MODE
CHANGE
Normal
-> Pitch
LASER ON
UTILITY
NOR
MAL
P. 1
AREA CHECK
INITIALIZE
ON
PANEL
mm
ONE-SCAN MODE
PROGRAM SELECT
DISPLAY UNIT
ESC
Program selection
Panel
Terminal
Front panel
Enabled
Disabled
Terminal input
Disabled
Enabled
RS-232C input
Enabled
Disabled
Program selection
Panel
Terminal
Front panel
Disabled
Disabled
Terminal input
Disabled
Enabled
RS-232C input
Enabled
Disabled