3 Function Settings
Setting the Mutual Interference Prevention Function
When two heads are connected and a little distance is left between them, a measurement error
may occur by receiving the light emitted from the other head. The mutual interference
prevention function is used in such a case.
When the mutual interference prevention function is set to ON, the two heads emit the laser
beam alternately to evade the interference from the other.
■
Setting procedure of the mutual interference prevention function
1
Hold down the
key, and select "
oPt on
" with the
key.
Set the common settings.
2
Press the
key, and select function
"
b
" with the
key.
Set the mutual interference prevention function.
3
Press the
key, and select
"
b-1
" with the
key.
The mutual interference prevention function is set to
"ON" in
this example.
4
Press the
key to register the setting. Press the
key to return to the measurement state.
The sampling rate is doubled when the mutual interference prevention function is set to
ON.