Setting the unit and the minimum display unit, Setting the unit and the minimum, Display unit -33 – KEYENCE LK-G Series User Manual
Page 91

3-33
3
LK-G-M-NO3-E
3 Function Settings
Setting the Unit and the Minimum Display Unit
The display unit and the minimum display unit can be selected from the following six types of
units.
The factory default setting is “G-2”. When using LK-G405/G400/G407/G402/G505/G500/G507/
G502, select function No. G-1 or G-0.
■
Setting procedure of the number of digits to be displayed and the position of the
decimal point
1
Hold down the
key, and select "
oUt-1
" with the
key.
OUT 1 is set in this example.
2
Press the
key, and select function "
G
" with the
key.
Set the display unit and the minimum display unit.
3
Press the
key, and select "
G-3
" with the
key.
Set the unit to mm and the minimum display unit to 0.00001.
4
Press the
key to register the setting. Press the
key to return to the measurement state.
If the unit and the minimum display unit are changed, the setting content related to the
numerical value display within the same OUT in the same program No. and the auto-zero
setting value is initialized.
Function No.
Minimum display unit Unit
Displayable range
g-0
0.01
0.01
mm
– 9999.99 to + 9999.99
g-1
0.001
0.001
mm
– 999.999 to + 999.999
g-2
0.0001
0.0001
mm
– 99.9999 to + 99.9999
g-3
0.00001
0.00001
mm
– 9.99999 to + 9.99999
g-4
0.1
0.1
µm
– 99999.9 to + 99999.9
g-5
0.01
0.01
µm
– 9999.99 to + 9999.99
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM
Note