Measuring with offset, Measuring with offset -32 – KEYENCE LK-G Series User Manual
Page 90

3-32
3
LK-G-M-NO3-E
3 Function Settings
Measuring with Offset
You can add or subtract any value to or from the displayed value. When the offset value is set, it
can be displayed when the auto-zero is performed.
The offset value is set for the measurement value that underwent a measurement mode process
and the auto-zero process.
Adjusting the master workpiece by using the auto-zero function (auto offset function)
The size of the master workpiece can be set as the master value (reference value), if you
set the size of the master workpiece as the offset value, and then use the auto-zero func-
tion while measuring the master workpiece.
■
Setting procedure of the offset value
The setting procedure is described below for the case where offset is set for OUT1 as an
example.
1
Hold down the
key, and select "
oUt-1
" with the
key.
OUT1 is chosen for this example.
2
Press the
key, and select function "
F
" with the
key.
Set the offset.
3
Press the
key, and set the offset value with the
and
keys.
"0.03" is set in this example.
4
Press the
key to register the setting. Press the
key to return to the measurement state.
Refer to "List of Default Values and Setting Ranges" (3-15 page) for the setting range and
its restriction.
Reference
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM
OUT1
HI
GO
LO
HI
GO
LO
TIM
OUT2
TIM